Ventilator simulator

Pick a patient, change the settings, and watch the waveforms, pressures, and blood gas respond. Each patient has a goal. Changes take effect on the next breath, like a real ventilator.

Pawcm H₂O
FlowL/min
VolumemL
PIP
–
Pplat
–
PEEP set
–
Driving P
–
VTe
–
VT / PBW
–
RR
–
VE
–
I:E
–
Mean Paw
–

Holds measure plateau pressure and total PEEP on the next breath.

How this simulator works

The lung is one compartment with a fixed resistance and compliance, on a sedated patient who doesn't trigger breaths. Airway pressure is PEEP + volume ÷ compliance + resistance × flow. Exhalation is passive, so a lung with a long time constant (resistance × compliance) can't empty before the next breath when the rate is fast: that trapped volume is auto-PEEP. In the ARDS patient, PEEP recruits collapsed lung, which raises compliance and lowers shunt until high PEEP starts to overdistend.

The blood gas is a steady-state estimate: PaCO2 = 0.863 × CO2 production ÷ alveolar ventilation, and PaO2 comes from mixing shunted and non-shunted blood through the oxyhemoglobin curve, so FiO2 helps little when the problem is shunt. Real patients differ. This is a study tool, not a clinical one. Related reading: ventilator graphics, initial settings, adjusting ventilation from ABGs, and ARDS.