Study tool
Ventilator simulator
Pick a patient, change the settings, and watch the waveforms, pressures, and blood gas respond. Each patient has a goal. Changes take effect on the next breath, like a real ventilator.
- PIP
- –
- Pplat
- –
- PEEP set
- –
- Driving P
- –
- VTe
- –
- VT / PBW
- –
- RR
- –
- VE
- –
- I:E
- –
- Mean Paw
- –
Holds measure plateau pressure and total PEEP on the next breath.
How this simulator works
The lung is one compartment with a fixed resistance and compliance, on a sedated patient who doesn't trigger breaths. Airway pressure is PEEP + volume ÷ compliance + resistance × flow. Exhalation is passive, so a lung with a long time constant (resistance × compliance) can't empty before the next breath when the rate is fast: that trapped volume is auto-PEEP. In the ARDS patient, PEEP recruits collapsed lung, which raises compliance and lowers shunt until high PEEP starts to overdistend.
The blood gas is a steady-state estimate: PaCO2 = 0.863 × CO2 production ÷ alveolar ventilation, and PaO2 comes from mixing shunted and non-shunted blood through the oxyhemoglobin curve, so FiO2 helps little when the problem is shunt. Real patients differ. This is a study tool, not a clinical one. Related reading: ventilator graphics, initial settings, adjusting ventilation from ABGs, and ARDS.